Heated Power-Plant-Probe with integrated heated filter in front of probe Ø 64 / 60 mm
The probe can be used in power plants with high concentration of sulphuric acid droplets in the flue gas.
- plane filter holder in the front part of the heated suction tube,
- separate controlled high performance heater cartridge around plane filter and conical entry,
- thermocouple measures the gas temperature to avoid condensation,
- no wrongness dust measurement take place.
Not in the delivery capacity: Bent, cone and plane filter holder!
See leaflet VGB-Powertech M 301.
Technical data:
- electrical heated 230 V with two separate heater cartridges,
- performance filter device: 400 W, with integrated Thermocouple NiCrNi,
- suction tube ca. 400 W/m, with integrated Thermocouple NiCrNi,
- integrated Thermocouple NiCrNi to measure the gas temperature behind the plane filter,
- support for the plane filter Ø 45 mm,
- fixed inner tube 10 x 1 mm,
- outer-Ø 65 mm, after ca. 150 mm outer-Ø 60 mm,
- two 220 V power supply and two connecttions for the thermocouple NiCr-Ni to regulate the front and probe heater,
- material: Titanium.